The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.
Source facts
Action
Philips Medical System sent an Customer Notification letter dated October 19, 2017, and plans to have a Philips Field Service Engineer go to site to perform the field change order (FCO70900041) including:
1. Check for the presence of the PE cable and, if missing, install it.
2 .Perform a modified screw fixation.
They expect to complete action by May 2018. For further questions, please call (978) 659-3000.
Initiated
2017-10-19
Root cause
Nonconforming Material/Component
Terminated
2018-09-20
Distribution
Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.
Recall number
Z-0330-2018
Recall status
Terminated
CDRH record ID
160799
FDA record created
2018-01-10
Recall event number
78880