Z-0331-2022: Dimension Vista 500- In-vitro diagnostic device intended to duplicate manual analytical procedures such a

Recall record Z-0331-2022 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0331-2022
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Dec 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays

Source facts

Action
Siemens issued Urgent Medical Device Correction letter dated 10/27/21 to US Customers (VSW-22-01.A.US) and OUS Customers will be sent Urgent Field Safety Notice (VSW-22-01.A.OUS) to inform them of the issue and instruct them on setting the HIL index for the affected assays if applicable. The Dimension Vista software has been updated with the correct HIL index default settings recommended in this letter and included in software version 3.10.2. Letter states reason for recall, health risk and action to take: For the products listed in Table 1, please perform the following steps if the HIL feature is activated on your Dimension Vista System. If your laboratory is not using the HIL feature and it is not activated on your system no action is required: If the HIL indices for Dimension Vista FOL are activated, be sure to set the I index (icterus) to 6 which is aligned with the information in the Dimension Vista FOL IFU. If the HIL indices for Dimension Vista AMM are activated, be sure to set the L index (lipemia) to 2 which is aligned with the information in the Dimension Vista AMM IFU. Note: Instructions for setting HIL indices can be found in the Dimension System Operator s Guide, in Section Configuration, System Configuration. If the HIL indices for Dimension Vista CTNI are activated, be sure to set the I index (icterus) to 7 which is aligned with the information in the Dimension Vista CTNI IFU. If the HIL indices for Dimension Vista TPSA and/or FPSA are activated, be sure to set the H index (hemolysis) to 7 and the I index (icterus) to 6 which is aligned with the information in the Dimension Vista TPSA and FPSA IFUs. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days to Siemens Healthineers for reporting to the Authorities. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your lo
Initiated
2021-10-15
Root cause
Software design
Terminated
2024-05-13
Distribution
Worldwide distribution - US Nationwide distribution and the countries of DE, ES, FR, IT, CA, NL, JP, PT, AU, CZ, AT, DK, SI, BE, SK, QA, SA, KR, CH.
Recall number
Z-0331-2022
Recall status
Terminated
CDRH record ID
190257
FDA record created
2021-12-02
Recall event number
89008

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗