Open original FDA record ↗Recall record
- FDA identifier
- Z-0331-2027
- Company in FDA record
- AVID Medical, Inc.
- FDA classification date
- Nov 3, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for open seals on header bags, compromising sterility.
Source facts
- Action
- AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to Segregate, Quarantine, and Inspect all kits identified. If the header bag seal appears to be compromised, discard the entire kit. Consignees were also requested to complete and return the provided Recall Response Form. Distributors were instructed to notify customers if affected kits were further distributed. On 10/06/2026 AVID Medical sent an updated "URGENT: MEDICAL DEVICE CORRECTION" letter to consignees via email. The strategy was updated and consignees were instructed not to use the affected kits and that they should either discard or return them. If returning, they should contact customer service for a Return Goods Authorization. Consignees were again requested to complete and return the provided Recall Response Form. Distributors were again instructed to notify customers if affected kits were further distributed.
- Initiated
- 2026-09-08
- Root cause
- Process control
- Distribution
- U.S Nationwide distributor in the state of Illinois.
- Recall number
- Z-0331-2027
- Recall status
- Open, Classified
- CDRH record ID
- 223012
- FDA classified
- 2026-11-03
- Recall event number
- 99967