Open original FDA record ↗Recall record
- FDA identifier
- Z-0332-2017
- Company in FDA record
- Spacelabs Healthcare Inc
- FDA classification date
- Nov 2, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.
Source facts
- Action
- The firm sent the Urgent - Medical Device Correction letter, dated 18 October 2016, via priority service, return receipt requested, to all U.S. customers on 10/18/16. The firm will e-mail a customer letter (translated as necessary) to all international subsidiaries and distributors of record on October 25, 2016. Xhibit Central Stations used without Xhibit Telemetry Receivers, are not affected. Customers were asked to circulate the letter to all person affected and place a copy with the appropriate monitor user manual until corrections have been implemented. Spacelabs Healthcare will contact affected consignees to schedule a convenient time for Spacelabs to update the affected Xhibit Central Station(s) with a new software version 1.1.6 at no cost. Customers with questions can call Technical Support at 1-800-522-7025 and select 2.
- Initiated
- 2016-10-18
- Root cause
- Software design
- Terminated
- 2017-04-03
- Distribution
- Worldwide Distribution -- USA, to AL, CA, CT, FL, GA, IA, ID, IL, IN, KY, MI, MN, MS, NC, NE, NJ, NY, OH, OK, PA, RI, SC, SD, TX, VA, WA, WY, and Puerto Rico; and, to the countries of AUSTRALIA, CANADA, CHILE, CZECH REPUBLIC, FRANCE, MEXICO, NETHERLANDS, PANAMA, SAUDI ARABIA, SWITZERLAND, TAIWAN, and UNITED KINGDOM.
- Recall number
- Z-0332-2017
- Recall status
- Terminated
- CDRH record ID
- 150467
- FDA classified
- 2016-11-02
- Recall event number
- 75477