Open original FDA record ↗Recall record
- FDA identifier
- Z-0332-2022
- Company in FDA record
- Covidien Llc
- FDA record creation date
- Dec 3, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
Source facts
- Action
- Medtronic customer communication translated by in-country representatives emailed on November 12, 2021. Letter states reason for recall, health risk and action to take: 1. Please immediately quarantine and discontinue use of the affected item code with the associated serial numbers listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated serial numbers must be returned. 3. If you have distributed the Covidien Sonicision" Battery Chargers listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. Fax to 800-895-6140 or email to rs.covidienfeedbackcustomerservice@medtronic.com Credit for returned affected product will be issued based on the RGA number. Exception: Customers with zero inventory, fax to 651-367-2624 or email to rs.gmbfcamitg@medtronic.com
- Initiated
- 2021-11-12
- Root cause
- Under Investigation by firm
- Distribution
- Brazil
- Recall number
- Z-0332-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190299
- FDA record created
- 2021-12-03
- Recall event number
- 89024