Open original FDA record ↗Recall record
- FDA identifier
- Z-0332-2027
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Nov 3, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Source facts
- Action
- Customers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, musculoskeletal injury and/or interruption of a surgical procedure. To date, Baxter has not received any reports of patient injury related to this issue. Baxter will call to schedule a time to come in and perform the belt tension verification. Customers can continue to use the tables.
- Initiated
- 2026-09-14
- Root cause
- Equipment maintenance
- Distribution
- US Nationwide distribution in the states of IL, MO, TX.
- Recall number
- Z-0332-2027
- Recall status
- Open, Classified
- CDRH record ID
- 223130
- FDA classified
- 2026-11-03
- Recall event number
- 99826