Z-0332-2027: TruSystem 7000 U Operating Table (dV), Product Code 1723633

Recall record Z-0332-2027 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0332-2027
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Nov 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

Source facts

Action
Customers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, musculoskeletal injury and/or interruption of a surgical procedure. To date, Baxter has not received any reports of patient injury related to this issue. Baxter will call to schedule a time to come in and perform the belt tension verification. Customers can continue to use the tables.
Initiated
2026-09-14
Root cause
Equipment maintenance
Distribution
US Nationwide distribution in the states of IL, MO, TX.
Recall number
Z-0332-2027
Recall status
Open, Classified
CDRH record ID
223130
FDA classified
2026-11-03
Recall event number
99826

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗