Open original FDA record ↗Recall record
- FDA identifier
- Z-0333-2022
- Company in FDA record
- Northgate Technologies, Inc.
- FDA record creation date
- Dec 3, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The languages for Russian, Romanian, Slovak, and Czech have the power settings for high and low switched in the device.
Source facts
- Action
- The recalling firm, Northgate Technologies Inc., issued a recall letter dated 10/25/2021 to their customer explaining the reason for recall. The customer was requested to determine the location of the affected devices. If the device is/could be used in one of the affected languages, the device is to be returned so the software can be updated to address the text mix-up. If the device is not or will not be operated in the affected languages, the device does not need to be returned immediately and the software will be upgraded during the next servicing to address the mix-up. Complete the enclosed Recall Acknowledgment Form indicating the serial numbers, distribution dates, and whether the device needs correction now or later and email it to: customerservice@ntisurgical.com. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product recall. If you have any questions, please contact me at (224) 856-2250 or by e-mail, tgatto@ntisurgical.com.
- Initiated
- 2021-10-25
- Root cause
- Software design (manufacturing process)
- Distribution
- US Distribution to state of MA.
- Recall number
- Z-0333-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190077
- FDA record created
- 2021-12-03
- Recall event number
- 88944