Skip to content

Public FDA record

Z-0334-2017: Guider Softip XF; GUIDER/MP XF/8FR/100 cm, Model number: M003101480. Cardiology: Boston Scientific Guid

Recall record Z-0334-2017 for Stryker Neurovascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0334-2017
Company in FDA record
Stryker Neurovascular
FDA classification date
Nov 2, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.

Source facts

Action
Stryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated October 5, 2016, to all affected customers by Federal Express. Letters advised of the reason for recall and models and devices affected and requested that all affected devices in inventory be returned to Stryker. Customers were advised to circulate the Field Safety Notice internally to all interested/affected parties and to complete the attached customer response form and return the form to their nominated Stryker Representative or to NVFieldActions@stryker.com. For questions regarding this recall call 510-413-2106.
Initiated
2016-10-05
Root cause
Employee error
Terminated
2017-03-23
Distribution
Worldwide Distribution - US including AK, IL, OH and Internationally to: ARGENTINA AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CHILE CZECH REPUBLIC ECUADOR EGYPT FRANCE GERMANY GREECE HUNGARY INDIA IRAN ISRAEL ITALIA JAPAN MEXICO NETHERLANDS NORWAY PERU PHILIPPINES POLAND RUSSIA SERBIA SINGAPORE SLOVAKIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND THAILAND TUNISIA TURKEY UK URUGUAY VIETNAM.
Recall number
Z-0334-2017
Recall status
Terminated
CDRH record ID
150386
FDA classified
2016-11-02
Recall event number
75447

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗