Skip to content

Public FDA record

Z-0334-2022: O3D ORCHESTRATE Orthodontic Technologies User Manual and Data Processing Instructions (model preparation

Recall record Z-0334-2022 for Orchestrate Orthodontic Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0334-2022
Company in FDA record
Orchestrate Orthodontic Technologies, Inc.
FDA record creation date
Dec 6, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to product label/labeling being altered from it's approved state.

Source facts

Action
On 11/11/2021, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" notification letter to customers informing them they discovered that the labeling sent to customers were not the approved version of the product labeling. The labeling was missing the following items; warnings, cautions, precautions, anticipated adverse events, indication for use, contraindications, software overview and software description. The incorrect labeling could result in the following risks/hazards/harms: - A lack of understanding of the Indications for Use of this software. - Use of the software by someone who is not a dental professional and does not have a thorough understanding of the dentistry and orthodontic principles, clinical applications, and risks. - The improper use of poor-quality dental scans as an input for treatment planning. - The lack of knowledge of the tools available for use during collision detection. - The sale of this product to someone other than a dental professional. - The user not knowing that they needed the necessary product training to use the software. - A lack of understand that a failure to observe all warnings and precautions may result in complications. - A lack of knowledge of theoretical adverse events that could result during sequential aligner treatment, after using this software. - A lack of knowledge regarding the technical description of the product. - A lack of knowledge regarding the technical requirements of the system. - Poor fit, possibly resulting in: ++ Poor fit ++ Infection ++ TMD ++ Crown Removal - Inappropriate tooth movements, including: ++ Nerve Damage ++ Pulpal Degradation ++ Gingival Recession ++ Root Resorption ++ Tooth Extrusion ++ Tooth Loss ++ Unintended tooth movement Customer are informed to: 1) Read the document (MKT-LB-005) that is included with the the customer notification. The document includes all information missing from the label sent with the delivery of the product. 2) Send a read receipt associated with the ema
Initiated
2021-11-11
Root cause
Process control
Terminated
2024-01-10
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. The countries of Argentina, Australia, Bolivia, Brazil, Canada, Costa Rica, India, Ireland, Mongolia, New Zealand, Pakistan, Spain and Thailand.
Recall number
Z-0334-2022
Recall status
Terminated
CDRH record ID
190266
FDA record created
2021-12-06
Recall event number
89013

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗