Skip to content

Public FDA record

Z-0335-2018: Intellicuff Standalone (PN 951001) Used to continuously measure and automatically maintain the user-set

Recall record Z-0335-2018 for Hamilton Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0335-2018
Company in FDA record
Hamilton Medical, Inc.
FDA record creation date
Jan 12, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Issue related to the performance of the motor in recalled product. During use, motor may cease to function.

Source facts

Action
Hamilton Medical sent a Recall Notice dated June 20, 2017, notifying them of the device correction. Consignees were informed that a Hamilton Representative would be in contact with them regarding the return of any devices. For further questions, please call (775) 858-3200.
Initiated
2017-06-20
Root cause
Component design/selection
Distribution
US Distribution to IL and TX.
Recall number
Z-0335-2018
Recall status
Open, Classified
CDRH record ID
156855
FDA record created
2018-01-12
Recall event number
77617

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗