Open original FDA record ↗Recall record
- FDA identifier
- Z-0335-2022
- Company in FDA record
- Osteomed, LLC
- FDA record creation date
- Dec 6, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to consoles not in compliance with the latest electrical safety standards IEC 60601 and as a result shocked a patient.
Source facts
- Action
- On 10/28/2021, the firm initiated a telephone communication and sent an "URGENT Medical Device Recall" communication/letter informing customers that the Power Consoles and Footswitches do not meet the current electrical and safety testing requirements for medical devices per IEC 60601 and as such, could result in possible electrical shocks to patients and users and potentially cause interference with other devices. Customer are instructed to: 1) Immediately discontinue use of devices by powering off, unplug the device and move the device to a separate room away from the point of care. 2) Review the Notification letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or fax within the 7 calendar days. Fax: 972.677.4766 Email: RecallNotificationGroup@acumed.net 3) Return devices within the scope of the recall and unused consumable inventory and 4) immediately discontinuing use of the devices within the scope of this voluntary recall. OsteoMed will provide FedEx shipping account information. In addition, we have issued a Return Material Authorization specific to your facility. This is located on the Acknowledgement and Receipt Form. 5) Distributors are being notified of this recall and involved in the communication. Distributors are requested to send their devices back with the RMA provided. The Recalling Firm is: initiated this voluntary recall and is requesting that all affected units be removed from point of care to prevent use of the device. The Recalling Firm will provide return information and point of care destruction should customers opt to dispose of the device at their facilities. For any questions, contact: Monday-Friday, 8 AM 5 PM CDT Kathryn A. Jayne Sr. Director, Quality Assurance &Regulatory Affairs OsteoMed LLC 972.677.4766 kathryn.jayne@acumed.net RecallNotificationGroup@acumed.net
- Initiated
- 2021-10-28
- Root cause
- Device Design
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, Guam, HI, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA and WI. The countries of Argentina, Bolivia, Brazil, Canada, Chile, Colombia, Dominican Republic, Ecuador, Germany, Guatemala, Indonesia, Italy, Japan, Korea, Kuwait, Lithuania, Malaysia, Mexico, Russia, Spain, Thailand, Turkey, United Kingdom, and Venezuela.
- Recall number
- Z-0335-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190208
- FDA record created
- 2021-12-06
- Recall event number
- 88990