Skip to content

Public FDA record

Z-0335-2022: OsteoMed OsteoPower Power Console: -OsteoMed OsteoPower Power Console 2 REF 450-0021-00 -OsteoMed D-Powe

Recall record Z-0335-2022 for Osteomed, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0335-2022
Company in FDA record
Osteomed, LLC
FDA record creation date
Dec 6, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to consoles not in compliance with the latest electrical safety standards IEC 60601 and as a result shocked a patient.

Source facts

Action
On 10/28/2021, the firm initiated a telephone communication and sent an "URGENT Medical Device Recall" communication/letter informing customers that the Power Consoles and Footswitches do not meet the current electrical and safety testing requirements for medical devices per IEC 60601 and as such, could result in possible electrical shocks to patients and users and potentially cause interference with other devices. Customer are instructed to: 1) Immediately discontinue use of devices by powering off, unplug the device and move the device to a separate room away from the point of care. 2) Review the Notification letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or fax within the 7 calendar days. Fax: 972.677.4766 Email: RecallNotificationGroup@acumed.net 3) Return devices within the scope of the recall and unused consumable inventory and 4) immediately discontinuing use of the devices within the scope of this voluntary recall. OsteoMed will provide FedEx shipping account information. In addition, we have issued a Return Material Authorization specific to your facility. This is located on the Acknowledgement and Receipt Form. 5) Distributors are being notified of this recall and involved in the communication. Distributors are requested to send their devices back with the RMA provided. The Recalling Firm is: initiated this voluntary recall and is requesting that all affected units be removed from point of care to prevent use of the device. The Recalling Firm will provide return information and point of care destruction should customers opt to dispose of the device at their facilities. For any questions, contact: Monday-Friday, 8 AM 5 PM CDT Kathryn A. Jayne Sr. Director, Quality Assurance &Regulatory Affairs OsteoMed LLC 972.677.4766 kathryn.jayne@acumed.net RecallNotificationGroup@acumed.net
Initiated
2021-10-28
Root cause
Device Design
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, Guam, HI, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA and WI. The countries of Argentina, Bolivia, Brazil, Canada, Chile, Colombia, Dominican Republic, Ecuador, Germany, Guatemala, Indonesia, Italy, Japan, Korea, Kuwait, Lithuania, Malaysia, Mexico, Russia, Spain, Thailand, Turkey, United Kingdom, and Venezuela.
Recall number
Z-0335-2022
Recall status
Open, Classified
CDRH record ID
190208
FDA record created
2021-12-06
Recall event number
88990

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗