Open original FDA record ↗Recall record
- FDA identifier
- Z-0336-2011
- Company in FDA record
- Ethicon Endo-Surgery Inc
- FDA record creation date
- Jun 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This recall has been initiated because Ethicon Endo-Surgery is aware of an increase in reports of leaking in the ENDOPATH XCEL" with OPTIVIEW" Technology trocar, which potentially can lead to loss of pneumoperitoneum during a surgical procedure. Pneumoperitoneum leaks may lead to conversion from laparoscopic to open surgery.
Source facts
- Action
- Ethicon Endo-Surgery sent an URGENT: DEVICE RECALL EVENT 2244 letter, dated October 8, 2010. The letter identified the product, the problem, and the action to be taken by the customer. The customers were instructed to stop using the product, examine all inventory to determine if they had the affected product on hand, and remove the affected products from inventory. Ethicon Endo-Surgery requested that each customer complete the Business Reply Form and fax it back to Stericycle at (877) 272-6141 with in 3 business days and also return any affected product to Stericycle. For question regarding this recall call (877) 598-5706 and reference 2244.
- Initiated
- 2010-09-27
- Root cause
- Device Design
- Terminated
- 2012-01-25
- Distribution
- Worldwide distribution: USA, ARGENTINA, AUSTRALIA, BELGIUM,BRAZIL,CANADA,CHILE,CHINA,COLOMBIA,COSTA RICA, DOMINICAN REPUBLIC,ECUADOR, EL SALVADOR, GREECE, GUATEMALA, HONG KONG, ISRAEL, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, LEBANON, NETHERLANDS ANTILLES, NICARAGUA, PANAMA SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA SPAIN, TRINIDAD AND TOBAGO, UNITED ARAB EMIRATES URUGUAY, and VENEZUELA.
- Recall number
- Z-0336-2011
- Recall status
- Terminated
- CDRH record ID
- 95076
- FDA record created
- 2015-06-26
- Recall event number
- 56939