Open original FDA record ↗Recall record
- FDA identifier
- Z-0337-2013
- Company in FDA record
- Superstat Corp
- FDA classification date
- Nov 16, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The recall was initiated because Superstat Corporation has confirmed that no documentation are available to verify that quarterly dose sterility audits were performed for the Superstat Modified Collagen Hemostatic Sponge.
Source facts
- Action
- Superstat initially contacted customers by telephone and sent an Urgent - Device Recall Notification Letter on October 24, 2012, to all affected customers. The recall letter provided the reason for recall was due to inadequate documentation of quarterly dose sterility audits. The recall includes all lots and part numbers of Superstat produced between 2002 and 2012. Customers are instructed to complete and return the enclosed response form as soon as possible. Customers with questions are instructed to call customer service at (800) 487-3786. For questions regarding this recall call 310-605-1655.
- Initiated
- 2012-10-24
- Root cause
- Process control
- Terminated
- 2013-02-06
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, Thailand, Mexico, Canada, Pakistan, Israel, Greece, Serbia, Canda, Isle of Man, Nova Scotia, and Singapore.
- Recall number
- Z-0337-2013
- Recall status
- Terminated
- CDRH record ID
- 114273
- FDA classified
- 2012-11-16
- Recall event number
- 63581