Open original FDA record ↗Recall record
- FDA identifier
- Z-0337-2014
- Company in FDA record
- GE Healthcare It
- FDA classification date
- Nov 18, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
1) Under rare network conditions, a Centricity Perinatal process may maintain an incorrect count of Patient slots in use which may lead to an inability to access the patient roster. The user is temporarily unable to review or edit form and chart information. This could result in a potential delay of care; however active fetal strips and permanent patient data storage are not affected. 2) When
Source facts
- Action
- GE Healthcare sent an URGENT MEDICAL DEVICE CORRECTION letter dated March 11, 2013, to all affected customers. The letters requested that direct accounts ensure that all potential users of the software were made aware of the potential problems and the recommended actions. Direct accounts were asked to return a customer response form and to return it to GE Healthcare via fax at 847-939-1522 or e-mail at QARAFMl1@ge.com. Version 6.94 of the software has been modified to correct these defects. GE Technical Support: 800-433-2009.
- Initiated
- 2013-03-11
- Root cause
- Software design
- Terminated
- 2016-06-10
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Australia, Canada, Denmark, Germany, Hong Kong, Korea, South Korea, Thailand and United Kingdom.
- Recall number
- Z-0337-2014
- Recall status
- Terminated
- CDRH record ID
- 117104
- FDA classified
- 2013-11-18
- Recall event number
- 64805