Open original FDA record ↗Recall record
- FDA identifier
- Z-0337-2016
- Company in FDA record
- NxStage Medical, Inc.
- FDA record creation date
- Nov 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed
Source facts
- Action
- NxStage issued a Field Safety Notice (FSN), dated October 29, 2015, to clinic and home patients with a NX1000-3 or NX1000-4 cycler. For patients performing short daily hemodialysis, the FSN included instructions for steps to be taken so that the cycler and treatments are not affected by the software error. Patients performing nocturnal hemodialysis were instructed to contact their Center for a replacement NX1000-1 Cycler. A software update will be released to correct the error. Customers should complete and return the reply form. Contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243) with questions or comments.
- Initiated
- 2015-10-29
- Root cause
- Software design
- Terminated
- 2019-11-19
- Distribution
- Worldwide Distribution.
- Recall number
- Z-0337-2016
- Recall status
- Terminated
- CDRH record ID
- 141528
- FDA record created
- 2015-11-25
- Recall event number
- 72554