Z-0338-2010: Hill-Rom 100 Low Bed, Model P3930. AC-powered adjustable hospital bed.

Recall record Z-0338-2010 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0338-2010
Company in FDA record
Hill-Rom, Inc.
FDA classification date
Nov 18, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The siderails may exhibit false latching, or unintentional lowering, or may become inoperable.

Source facts

Action
Hill Rom notified consignees by recall letter dated 4/20/09 and advised them of the problem and requested them to locate the beds. Consignees with patients in need of siderails were directed to be moved to a different model bed with siderails and, if the patient did not need siderails; consignees were requested to remove the siderails, move the siderails to a stowed position, to tie the siderail to the bed frame or to tag the siderail as out of service. The letter stated the consignee would be contacted within 90 days by a firm representative to do a correction on all affected units.
Initiated
2009-04-20
Root cause
Process control
Terminated
2010-10-15
Distribution
Worldwide distribution: USA, Canada, France, India, Singapore and Spain.
Recall number
Z-0338-2010
Recall status
Terminated
CDRH record ID
80169
FDA classified
2009-11-18
Recall event number
51389

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗