Open original FDA record ↗Recall record
- FDA identifier
- Z-0338-2011
- Company in FDA record
- Abbott Vascular
- FDA classification date
- Nov 10, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because Abbott vascular has discovered through internal testing that the affected lots may not meet the firm's quality specifications for catheter shaft tensile strength. If the affected device is used it may result in acute stent deployment difficulties and subsequent intervention.
Source facts
- Action
- Abbott Vascular sent an URGENT DEVICE RECALL letter dated September 10, 2010, to all consignees. The letter identified the product, the problem, and the action to be taken. The Sales Representatives were to personally visit each account to deliver the recall letter and assist in identifying and returning any unused devices to Abbott Vascular. Also, Abbott Vascular Regulatory Compliance initiated contact with the Risk Management department at each account by phone to verify their mailing address and provide a courtesy copy of the recall letter. The consignees were instructed to work with their local account representative to review their inventory, complete the Recall Effectiveness Check and return identified products to Abbott Vascular. For questions regarding this recall call (951) 914-3324.
- Initiated
- 2010-09-10
- Root cause
- Pending
- Terminated
- 2010-11-16
- Distribution
- Nationwide Distribution including AL, CA, CT, DC, FL, GA, IA, IL, IN, MA, ME, MI, MN, MO, OH, OK PA, TX, VA, and WI
- Recall number
- Z-0338-2011
- Recall status
- Terminated
- CDRH record ID
- 95084
- FDA classified
- 2010-11-10
- Recall event number
- 56973