Open original FDA record ↗Recall record
- FDA identifier
- Z-0338-2016
- Company in FDA record
- Masimo Corporation
- FDA classification date
- Nov 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
These sensors were manufactured with incompatible configurations. This could result in sensors that will provide either no readings or inaccurate readings for the SpCO and SpMet parameters.
Source facts
- Action
- Masimo sent an Urgent Product Recall letter on October 28, 2015, to all affected customers to inform them that Masimo has identified specific lots of sensors that were manufactured with incompatiable configurations. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete the tracking/verification form and email to Masimo. Customers with questions were instructed to contact Masimo Technical Services at http://Service.Masimo.com. For questions regarding this recall call 949-297-7000.
- Initiated
- 2015-10-28
- Root cause
- Finished device change control
- Terminated
- 2016-05-18
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Australia, Costa Rica, France, Germany, Kuwait, Martinique, Netherlands, Norway, South Korea, and Turkey.
- Recall number
- Z-0338-2016
- Recall status
- Terminated
- CDRH record ID
- 141684
- FDA classified
- 2015-11-25
- Recall event number
- 72626