Erroneous results issue when MONOJECT blood collection tubes are used during analysis after tubes are uncapped and recapped then analysed in the closed vial mode. Results are lower than if cap had not been removed.
Source facts
Action
A letter issued to users informing them of the incompatability of the Coulter ACT diff 2 Analyzer with the use of Monoject blood collection tubes made by Tyco Healthcare Group/Kendall Healthcare. Customers choosing to use these tubes are instructed to uncap the tubes and analyze the samples,as open vial samples.
Initiated
2004-12-03
Root cause
Other
Terminated
2012-01-13
Distribution
Nationwide and Canada
Recall number
Z-0339-05
Recall status
Terminated
CDRH record ID
36170
FDA classified
2004-12-21
Recall event number
30587