Certain Model NS-P probes contained in the NeuroSensor Cerebral Blood Flow and Intracranial Pressure Monitoring Kit may allow an Intracranial Pressure (ICP) drift of up to 5 mmHg during the first 24 hours of monitoring ICP. This is outside the products labeling specification of 2 mmHg maximum ICP drift within the first 24 hours of monitoring.
Source facts
Action
Notify Customers and Sales Representatives who received the product by letter sent via Federal Express. Customers will be instructed that an Integra Sales Representative will contact them to arrange for return of affected product.
Initiated
2005-09-08
Root cause
Other
Terminated
2012-03-07
Distribution
Nationwide
Recall number
Z-0339-06
Recall status
Terminated
CDRH record ID
41902
FDA classified
2006-01-06
Recall event number
33451