Open original FDA record ↗Recall record
- FDA identifier
- Z-0339-2023
- Company in FDA record
- Caire, Inc.
- FDA classification date
- Dec 2, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
Source facts
- Action
- The firm issued an URGENT: MEDICAL DEVICE RECALL notice its consignees be email and telephone. The notice explained the issue and requested the removal of the affected products from the field and return to CAIRE. CAIRE will immediately replace the above serial numbers with new Liberator devices. Please contact CAIRE Customer Service via phone at 1-800-482-2473 or 770-721-7700 or via email at customerservice.usa@caireinc.com to arrange for the return and replacement of the above serial numbers.
- Initiated
- 2022-10-10
- Root cause
- Process control
- Terminated
- 2024-08-19
- Distribution
- US Nationwide - Worldwide Distribution: CA, NC, OH, AZ, and Canada, Chile, Colombia, and Germany
- Recall number
- Z-0339-2023
- Recall status
- Terminated
- CDRH record ID
- 196753
- FDA classified
- 2022-12-02
- Recall event number
- 90999