Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.
Source facts
Action
Siemens has issued an Urgent Medical Device Recall in the US and an Urgent Field Safety Notice Outside the US on September 24, 2015, instructing customers to discontinue use, discard reagent and request no charge replacement kits for these products.
Initiated
2015-09-24
Root cause
Under Investigation by firm
Terminated
2017-01-13
Distribution
Distributed in the states of AZ, NY, CA, IN, LA, MN, OH, MI, NM, and GA, and the countries of Mexico and Canada.
Recall number
Z-0340-2016
Recall status
Terminated
CDRH record ID
140899
FDA classified
2015-11-25
Recall event number
72392