Skip to content

Public FDA record

Z-0341-2007: Cyberonics VNS Therapy System, Model 250-''HAND HELD'' programming software v6.1, v6.1.7, v7.0 and v7.1.3

Recall record Z-0341-2007 for Cyberonics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0341-2007
Company in FDA record
Cyberonics, Inc
FDA classification date
Jan 25, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician

Source facts

Action
Firm sent safety alert to physicians on 11/14/06. Safety Alert informs physicians of the problem and provides instructions on how to avoid an event and correct an event should it occur.
Initiated
2006-10-17
Root cause
Other
Terminated
2009-10-02
Distribution
Worldwide Distribution ---- USA and countries of Argentina, Australia, Austria, Belgium, Canada, Cyprus, Czech Rep, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, S. Korea, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, and Venezuela.
Recall number
Z-0341-2007
Recall status
Terminated
CDRH record ID
49143
FDA classified
2007-01-25
Recall event number
36672

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗