Transcutaneous pCO2-electrodes were wrongly coded as: pO2 electrode on the hardware plug for the firm's TCM series monitors. This causes the monitor to interpret the electrode wrongly, causing it to fail all calibrations and never get ready for use.
Source facts
Action
Radiometer contacted the affected U.S. customer and informed them of the recall via telephone on September 27, 2011. Radiometer requested the return of the affected product received from the firm. For questions call (440) 871-0463.
Initiated
2011-09-27
Root cause
Employee error
Terminated
2012-02-19
Distribution
Worldwide Distribution - USA (nationwide) in the state of San Diego, CA.
Recall number
Z-0341-2012
Recall status
Terminated
CDRH record ID
104616
FDA classified
2011-12-01
Recall event number
60153