Open original FDA record ↗Recall record
- FDA identifier
- Z-0341-2014
- Company in FDA record
- Synthes USA HQ, Inc.
- FDA classification date
- Nov 18, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The potential exists for the locking screw nut to break on the Soft Tissue Retractor if it is overt-tightened.
Source facts
- Action
- Synthes sent an Urgent Notice: Medical Device Recall letter dated July 31, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and immediately remove the affected lots from stock, call Synthes at 1-800-479-6329 to obtain a Return Authorization Number. Complete the Verification Section at the letter by checking the appropriate box indicating affected product has been located and return the Verification Form with the product to Credit/Returns, Synthes, 1101 Synthes Avenue, Monument, CO 80132. If customers do not have the identified product, they were instructed to complete the Offset blade facilitates easy preparation of the epipereosteal cavity for percutaneous plate insertion.e attached Verification Section and return to Synthes.
- Initiated
- 2013-09-24
- Root cause
- Device Design
- Terminated
- 2015-06-01
- Distribution
- Worldwide Distribution - USA (Nationwide) and Canada.
- Recall number
- Z-0341-2014
- Recall status
- Terminated
- CDRH record ID
- 122552
- FDA classified
- 2013-11-18
- Recall event number
- 66529