Open original FDA record ↗Recall record
- FDA identifier
- Z-0342-2012
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Dec 1, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.
Source facts
- Action
- The firm, Elekta, issued a "FIELD CHANGE ORDER MANDATORY ACTION" notice dated October 20, 2011 to its customers to add fuse protection for the kV transformers. The notice describes the product, problem, provides instructions and procedures and actions to be taken. The customers were instructed to follow any instructions or recommendations covered in the notice and to file the document in the Important Notice section of the appropriate User Manual. Elekta will contact the customer if the FCO is applicable to their kV generator. If you have any questions, call 770-300-9725.
- Initiated
- 2011-10-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-08-09
- Distribution
- Nationwide distribution: USA including states of: AL, AK, AZ, CA, CT, FL, GA, IL, IA, KY, MD, MS, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TX, VT, VA, WA, WV, WI, Puerto Rico, & DC.
- Recall number
- Z-0342-2012
- Recall status
- Terminated
- CDRH record ID
- 105583
- FDA classified
- 2011-12-01
- Recall event number
- 60497