Z-0342-2014: TruSystem Surgical Table (TS7000) Leg Section One Part Light HU Use of the Leg Section One Part Light H

Recall record Z-0342-2014 for Trumpf Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0342-2014
Company in FDA record
Trumpf Medical Systems, Inc.
FDA classification date
Nov 18, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The coupler point (hood) of the Leg Section One Part Light may become detached from the leg section frame.

Source facts

Action
TRUMPF sent an Urgent Safety Information letter dated October 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to inspect Leg Section One Part Light for damage to the coupler point (hook) between each case. If any damage is noticed to the coupler, they were instructed to stop use immediately and contact TRUMPF Technical Service. If they drop or collide the Leg Section One Part Light into anything then stop use immediately and contact TRUMPF Technical Service. Review the User Instruction for proper use and care of the device. The leg section should be used only as intended as part of the TRUMPF TS7000 surgical table. The intended use is to support and position the patients legs during the pre-operative and post-operative stages of surgery. The leg section should be removed during surgery. TRUMPF will be exchanging the hooks, counter plates and screws of the coupler point to increase the area that the load is distributed over. This will allow the screws securing the hood to the leg frame to withstand external forces. TRUMPF anticipates parts to be available the first week of November. TRUMPF trained technicians will complete the TRUMPF Medical Systems, Inc. for the part exchange. Completion is anticipated by November 11, 2013. Further questions please call (843) 534-0606
Initiated
2013-10-28
Root cause
Packaging
Terminated
2014-09-16
Distribution
US Distribution including the states of CA, IL, NJ, NY, NC, TX, WA and SC.
Recall number
Z-0342-2014
Recall status
Terminated
CDRH record ID
122967
FDA classified
2013-11-18
Recall event number
66688

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗