Z-0342-2020: Grab n Go Opti series VIPR system Model # PRX-9646 - Product Usage: The Grab n Go Opti system is an inte

Recall record Z-0342-2020 for Western / Scott Fetzer Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0342-2020
Company in FDA record
Western / Scott Fetzer Company
FDA record creation date
Nov 6, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Limited access to flow settings as a result of the control knob having been rotated beyond its functional range

Source facts

Action
Western formally communicated the product issue to Praxair on 7/12/19, who had been previously in contact with Western regarding this issue. Western plans to issue a formal communication to affected distributors in the near future. The firm has initiated corrective action in the field as of August 2019. The firm has implemented a new screening process at each fill plant to identify and isolate units that could potentially be rotated past the stop. Units which fail are isolated and returned to Western, which will inspect each failed unit, replace the failed component with the revised knob/plate, and return the refurbished unit to the fill plant. One customer specifically requested that, instead of the screening procedure above, all of its units in service be replaced with the revised design. Western fulfilled this request. On October 11, 2019, Western mailed a Medical Device Recall letter to its distributors to officially communicate the issue.
Initiated
2019-08-15
Root cause
Device Design
Terminated
2020-11-18
Distribution
Worldwide distribution US Nationwide distributions in the states of OH, NY, PA, TX, MA, MI, KS, IL, MD, MN, WA, TN, UT, CA, NJ, LA, CT, OR, ND, RI, WI, PR, FL, ON, NV, NB, OK, IN, AZ, MS, GA, IA and countries of Costa Rica, Mexico.
Recall number
Z-0342-2020
Recall status
Terminated
CDRH record ID
176512
FDA record created
2019-11-06
Recall event number
83857

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗