Open original FDA record ↗Recall record
- FDA identifier
- Z-0343-2016
- Company in FDA record
- Ormco/Sybronendo
- FDA classification date
- Nov 30, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ormco Corporation is recalling Copper NiTi Preformed Archwires because the product transformation temperature as listed on the labeling is incorrect.
Source facts
- Action
- A recall letter dated 11/5/15 was sent to all their customers who purchased the Copper NiTi Preformed Archwires because it has been brought to Ormco's attention that the product transformation temperature as listed on the labeling is incorrect. The wire is labeled as having 40 degrees Celsius transformation temperature, but the actual wire has a transformation temperature of 29 degrees Celsius. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete and fax back the enclosed acknowledgement form.
- Initiated
- 2015-11-04
- Root cause
- Employee error
- Terminated
- 2016-01-19
- Distribution
- Distributed US (nationwide) and the countries of Canada, Japan, Mexico, Russia, and Peru.
- Recall number
- Z-0343-2016
- Recall status
- Terminated
- CDRH record ID
- 141617
- FDA classified
- 2015-11-30
- Recall event number
- 72590