Skip to content

Public FDA record

Z-0344-2023: MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software ve

Recall record Z-0344-2023 for Mobius Imaging, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0344-2023
Company in FDA record
Mobius Imaging, LLC
FDA classification date
Dec 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.

Source facts

Action
On November 18, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. Customers were informed that a service representative will contact their facility to schedule a time to replace the batteries in the affected devices. The customer does not need to take additional action while waiting for the battery preplacement. If you have any questions or concerns regarding this update, please contact shirleytechnicalsupport@stryker.com or call Stryker Technical Support at 978 386 9619.
Initiated
2022-11-18
Root cause
Other
Terminated
2025-02-26
Distribution
Domestic distribution to consignees in Colorado, Tennessee, and Utah.
Recall number
Z-0344-2023
Recall status
Terminated
CDRH record ID
197190
FDA classified
2022-12-06
Recall event number
91178

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗