During treatments, using the Prostatron units, the devices' Temperature Sensing System may change the rectal temperature reading by up to 2 degrees Celsius and may briefly delay updates to the rectal temperature reading.
Source facts
Action
A letter about the problem, actions to take to avoid the problem, and the proposed software correction was sent to consignees on December 10, 2003. A software upgrade will be performed on all units in March 2004.
Initiated
2003-12-10
Root cause
Other
Terminated
2005-11-15
Distribution
The product was shipped to consignees located nationwide in the United States and to foreign consignees in Australia, France and Japan.
Recall number
Z-0345-04
Recall status
Terminated
CDRH record ID
30843
FDA classified
2004-07-20
Recall event number
27983