Open original FDA record ↗Recall record
- FDA identifier
- Z-0345-2013
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Nov 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies.
Source facts
- Action
- Beckman Coulter sent a field modification was released in August 2011 which included: - Installing a new aspirate and dispense tubing routing bracket. - Replacing existing aspirate and dispense probe tubing with modified tubing to help eliminatemisrouting of aspirate and dispense probe tubing. During installation the FSE informed the customer: -That aspirate and dispense tubing are different lengths. - That probes are not interchangeable with other instruments once tubing has been modified. - To route probe tubing by referring to the labels on the tubing. An updated service document, Routine Maintenance Changes, A82294B was provided to every customer. For questions regarding this recall call 714-961-4483.
- Initiated
- 2011-08-01
- Root cause
- Process design
- Terminated
- 2012-11-19
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, AE, AU, BG, BW, CH, CN, CZ, DE, ES, FR, GB, HK, HR, HU, IN, IT, JO, JP, LB, MA, MX, NL, PL, QA, RO, RU, SK, TN, TR, ZA, AO, AR, AT, BE, CO, DZ, FI, IE, IL, KR, KW, LI, NA, NO, NZ, PF, PH, PR, SE, SG, SZ, TH, TW, VE, VN, and YT.
- Recall number
- Z-0345-2013
- Recall status
- Terminated
- CDRH record ID
- 114074
- FDA classified
- 2012-11-19
- Recall event number
- 63505