DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper port, this defect results in the inability to read or display temperatures on the infant warming system.
Source facts
Action
DeRoyal issued their recall on 10/14/2015 to the end-user level by sending overnight letters to its direct accounts and hospitals. The consignees were instructed to destroy and discard the units. In the case of further distribution, the firm requested that the consignee either notify its customers of the recall or to submit a customer list for DeRoyal to contact them.
Initiated
2015-10-14
Root cause
Process control
Terminated
2017-07-01
Distribution
Distributed in the states of VA, AL, FL, and CA.
Recall number
Z-0345-2016
Recall status
Terminated
CDRH record ID
141078
FDA classified
2015-11-30
Recall event number
72454