This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.
Source facts
Action
The affected device was retrieved from the only customer on 9/4/19. An Urgent Medical Device Recall notification letter dated 9/12/19 was sent to that customer as well.
Initiated
2019-09-04
Root cause
Under Investigation by firm
Terminated
2020-01-21
Distribution
The product was distributed to GA
Recall number
Z-0345-2020
Recall status
Terminated
CDRH record ID
177019
FDA record created
2019-11-07
Recall event number
84025