Skip to content

Public FDA record

Z-0346-2023: Accelerate PhenoTest BC kit REF 10102018

Recall record Z-0346-2023 for Accelerate Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0346-2023
Company in FDA record
Accelerate Diagnostics Inc
FDA classification date
Dec 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.

Source facts

Action
On 11/04/2022, the firm communicated an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that the firm has become aware of complaints concerning false positive identification results for Staphylococcus aureus and Streptococcus spp. when using the Accerlerate PhenoTest BC kits. Customer are instructed to: (1) Share this customer notification with all who need to be made aware within their organization. (2) Appropriately destroy all affected Accelerate PhenoTest BC kits by defacing all labeling and disposing in accordance with site biohazard or waste policies and procedure. (3) Complete the Recall Response Form and return it to Accelerate Diagnostics. (4) Retrospectively review runs from affected lots, specifically those runs with positive identification results for either Staphylococcus aureus or Streptococcus spp. Follow their institutional policies and processes to review prior patient results related to the described issue. (5) Continue to report all technical issues through Accelerate Technical Support. Contact for Questions: NA Customer Service & Technical Support Toll free: +1 (888) 586 2939 opt. 4 Direct: +1 (520) 365 3100 Email: support@axdx.com
Initiated
2022-10-17
Root cause
Mixed-up of materials/components
Distribution
U.S.: AR, CA, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, MS, MT, NC, ND, NJ, NY, OH, PA, RI, SC, TN, TX, VA, WA, and WV. O.U.S.: Germany, Italy, Kuwait, Latvia, Poland, Portugal, Romania, Saudi Arabia, Spain, and United Arab Emirates
Recall number
Z-0346-2023
Recall status
Open, Classified
CDRH record ID
196684
FDA classified
2022-12-06
Recall event number
91075

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗