Z-0347-06: System 1000 family of Hemodialysis Instruments, including the System 1000, TINA, AURORA and ARENA; Baxter

Recall record Z-0347-06 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0347-06
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Jan 6, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition.

Source facts

Action
Baxter sent Urgent Device Correction letters dated 11/18/05 to all System 1000, TINA, Aurora and Arena Hemodilaysis Instrument customers via first class mail on the same date, to the attention of the Hemodialysis Administrator. The letters informed the accounts that the potential exists for fluid or air to be passed through the venouus line clamp that could result in an air emboli condition. This condition can occur when the tubing is not centered on the clamp and extends beyond the edge of the clamp''s pinch zone. Baxter will make arrangements to upgrade the instruments with either a linear or rotary venous line clamp upgrade as soons as they become available within the next three months. The letters included copies of a diagram showing the correct and incorrect placement of the tubing in the clamp, and a Technical Service Bulletin that provided an adjustment procedure to perform an interim correction to the linear line clamp configuration prior to the upgrade. Any questions regarding the execution of this procedure were directed to Baxter Global Technical Services at 1-800-553-6898.
Initiated
2005-11-18
Root cause
Other
Terminated
2008-03-04
Distribution
Nationwide and internationally to Canada, Korea, Hong Kong, Shanghai, Jamaica, Philippines, United Kingdom, Denmark, India, Belgium, Mexico, Italy, Austria, Sweden, Ecuador, Argentina, Brazil, Chile, Colombia, Peru, Turkey and Australia.
Recall number
Z-0347-06
Recall status
Terminated
CDRH record ID
42984
FDA classified
2006-01-06
Recall event number
34043

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗