The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
Source facts
Action
On September 30, 2026 Shimadzu Corp. issued a Urgent Medical Device Recall Notification to affected consignees via email. Shimadzu asked consignees to take the following actions;
1. If the bolts become loose, an abnormal noise will occur when moving the C-arm. Please check for abnormal noises during pre-study inspections.
2. If an abnormal noise is heard, please stop using the product and contact our service department.
3. Please complete the Medical Device Recall Customer Acknowledgement.
Initiated
2025-09-30
Root cause
Process control
Distribution
US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.
Recall number
Z-0347-2027
Recall status
Open, Classified
CDRH record ID
223189
FDA classified
2026-11-03
Recall event number
100044