Unintended single pulse maybe emitted to handpiece prior to pressing of trigger or footswitch and cause injury.
Source facts
Action
Candela notiifed Distributors on 11/8/13 through e-mail, Certified return receipt request, or FedEx package containing distributor letter, customer letter for translation and customer lists. A software update to Version 2.0 will be issued to the user by a field service representative.
Initiated
2013-11-06
Root cause
Software design
Terminated
2015-06-08
Distribution
Worldwide distribution: US (nationwide) and Internationally to: Azerbaijan, Bulgaria, Chile, Czech Republic, Greece, Israel, Kuwait, Qatar, Saudi Arabia, Serbia, Turkey, and the United Arab Emirates and to Subsidiaries in Australia, France, Germany, Italy, Japan, Portugal, Spain, and the United Kingdom.
Recall number
Z-0348-2014
Recall status
Terminated
CDRH record ID
123370
FDA classified
2013-11-18
Recall event number
66773
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.