Skip to content

Public FDA record

Z-0348-2014: Candela Laser GentleLase Pro, Powered Laser Surgical Instrument for dermatological use - Model number: 99

Recall record Z-0348-2014 for Candela Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0348-2014
Company in FDA record
Candela Corporation
FDA classification date
Nov 18, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unintended single pulse maybe emitted to handpiece prior to pressing of trigger or footswitch and cause injury.

Source facts

Action
Candela notiifed Distributors on 11/8/13 through e-mail, Certified return receipt request, or FedEx package containing distributor letter, customer letter for translation and customer lists. A software update to Version 2.0 will be issued to the user by a field service representative.
Initiated
2013-11-06
Root cause
Software design
Terminated
2015-06-08
Distribution
Worldwide distribution: US (nationwide) and Internationally to: Azerbaijan, Bulgaria, Chile, Czech Republic, Greece, Israel, Kuwait, Qatar, Saudi Arabia, Serbia, Turkey, and the United Arab Emirates and to Subsidiaries in Australia, France, Germany, Italy, Japan, Portugal, Spain, and the United Kingdom.
Recall number
Z-0348-2014
Recall status
Terminated
CDRH record ID
123370
FDA classified
2013-11-18
Recall event number
66773

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗