There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
Source facts
Action
Maquet issued an Urgent Medical Device Recall letter/Return Response form dated September 27, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customer were instructed to complete the return response form via fax to 1-973-629-1518. For questions contact Maquet Customer Service at 1-888-627-8383 option 2 followed by option 2 Monday through Friday between 8 amd and 6 pm. EST.
Initiated
2016-09-27
Root cause
Device Design
Terminated
2019-08-19
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-0348-2017
Recall status
Terminated
CDRH record ID
149889
FDA record created
2016-11-03
Recall event number
75271