Open original FDA record ↗Recall record
- FDA identifier
- Z-0349-2018
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA record creation date
- Jan 14, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Finishing guide does not adequately account for shape differences present in Size 8 of Persona Partial Knee System, potentially resulting in additional surgery time or sensation of tightness during knee flexion.
Source facts
- Action
- Customers were notified via FedEx mail and email on approximately 07/05/2017. Instructions included to notify appropriate personnel, complete and return the Acknowledgement Form, return affected product, and provide contact information of additional hospitals or surgeons that used the affected product.
- Initiated
- 2017-07-05
- Root cause
- Device Design
- Terminated
- 2019-04-02
- Distribution
- Distributed domestically to AZ, CA, CO, ID, IL, IN, MA, MD, MI, NJ, NM, PA, SC, TX, WA, WI. Distributed internationally to Belgium, France, Germany, Italy, Japan, Netherlands, Spain, Sweden, United Kingdom.
- Recall number
- Z-0349-2018
- Recall status
- Terminated
- CDRH record ID
- 160230
- FDA record created
- 2018-01-14
- Recall event number
- 78725