The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or cardioplegia sufficiently.
Source facts
Action
Letters were mailed to customers on 11/10/2006 by Federal Express.
Initiated
2006-11-10
Root cause
Other
Terminated
2007-02-05
Distribution
Hospitals in CA, NC, FL, WY, NY and WV.S
Recall number
Z-0350-2007
Recall status
Terminated
CDRH record ID
49779
FDA classified
2007-01-25
Recall event number
37023