Merge is conducting a hardware recall (replacement) of Fellows "Mighty 8" Power Strip and TrippLite Power Strip due to overheat of input resistor causing the power strip to smoke.
Source facts
Action
Merge began notify customers via an Urgent Medical Device Recall letter on 4/6/2017 via mail. The letter identified affected product, stated the issue, and informed customers that Merge would replace affected product. A response form is requested to be returned to Merge. Questions can be sent via email to recall@merge.com
Initiated
2017-04-06
Root cause
Component design/selection
Terminated
2019-07-10
Distribution
Nationwide.
Recall number
Z-0350-2018
Recall status
Terminated
CDRH record ID
154394
FDA record created
2018-01-14
Recall event number
76847