Open original FDA record ↗Recall record
- FDA identifier
- Z-0352-2024
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Nov 20, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.
Source facts
- Action
- An URGENT MEDICAL DEVICE CORRECTION notification letter dated 10/3/23 was sent to customers. Actions Required " Roche recommends using citrated plasma for D-Dimer quantification until further notice. " If your laboratory continues to use Li-heparin plasma for D-Dimer quantification, you are required to confirm all results between 0.5 g FEU/mL and 0.675 g FEU/mL with another D-Dimer method or with a citrated plasma sample. " If your laboratory continues to use Li-heparin plasma for D-Dimer quantification and chooses to confirm results with another method, you may request credit for that testing from Roche. Contact your Roche sales representative to request reimbursement. " Please complete all sections of the Recall Response Form (TP-01981) enclosed with this UMDC and fax it to 1-844-491-7776 or email it to roche5037@sedgwick.com. " File this Urgent Medical Device Correction (UMDC) for future reference. Credit for Confirmatory Testing If your laboratory continues to use Li-heparin plasma for D-Dimer quantification and chooses to confirm results with another method, you may request credit for that testing from Roche. Contact your Roche sales representative to request reimbursement. Distributor Actions Required " Read the enclosed Customer UMDC TP-01986. " Within 14 days from receipt of this UMDC, distribute a copy of the enclosed Customer UMDC TP-01986 to all customers to whom you have shipped the Tina-quant D-Dimer assays listed above in the last 2 years. " Please complete all sections of the Distributor Recall Response Form (TP-01985) enclosed with this UMDC and fax it to 1-844-491-7776 or email it to roche5037@sedgwick.com. " File this UMDC and the enclosed Customer UMDC for future reference. Questions Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
- Initiated
- 2023-10-03
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0352-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203829
- FDA classified
- 2023-11-20
- Recall event number
- 93282