Open original FDA record ↗Recall record
- FDA identifier
- Z-0353-2020
- Company in FDA record
- Heritage Medical Products, Inc.
- FDA record creation date
- Nov 8, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for bed to drop suddenly at the head while a patient is laying on it.
Source facts
- Action
- Heritage Medical Products, Inc. notified customers on about 09/02/2016 via "Advisory Notice." Customers were informed that Heritage Medical has designed a reinforcement kit as a correction, which will be supplied to customers. Customers were instructed to contact Heritage Medical, provide the serial number of their bed to ensure it is affected, and Heritage Medical will provide a free kit that will reinforce the area on the affected devices that has experienced failures. The contact information is a follows: Heritage Medical Products, Inc., ATTN.: Trunnion Reinforcement Kit, 10380 CR 6310, West Plains, MO 65775, (888) 256-8748.
- Initiated
- 2016-09-02
- Root cause
- Device Design
- Terminated
- 2020-12-14
- Distribution
- Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, OH, OK, OR, PA, SC, TX, UT, VA, WI, WY, PR. International distribution to Canada, Saudi Arabia, Trinidad & Tobago.
- Recall number
- Z-0353-2020
- Recall status
- Terminated
- CDRH record ID
- 174205
- FDA record created
- 2019-11-08
- Recall event number
- 83219