Skip to content

Public FDA record

Z-0353-2023: CytoCell BCL11B Proximal in FITC Spectrum-LDT for investigation T-cell related malignancy. This device is

Recall record Z-0353-2023 for Cytocell Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0353-2023
Company in FDA record
Cytocell Ltd.
FDA classification date
Dec 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device is mislabeled. The package insert for this ASR device contains a chromomap - an image showing the position the FISH probe maps to on the specific chromosome. For this device the chromomap in version 1 of the package insert is incorrect and doesn't not match the probe design or specification contained on the package insert

Source facts

Action
Cytocell issued Urgent Medical Device Recall letter on 10/27/22 to the end-user via email. Letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that all users review and implement the enclosed new package insert (v2) for this device. Please dispose of any copies of v1 of the package insert. Once implemented, all devices can continue to be used as the device itself has not malfunctioned. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Initiated
2022-10-27
Root cause
Under Investigation by firm
Terminated
2024-12-27
Distribution
TX
Recall number
Z-0353-2023
Recall status
Terminated
CDRH record ID
196769
FDA classified
2022-12-01
Recall event number
91117

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗