Z-0354-06: The PL V Continuum ventilator is a microprocessor controlled, compressorbased, mechanical ventilator. It

Recall record Z-0354-06 for Respironics California Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0354-06
Company in FDA record
Respironics California Inc
FDA classification date
Jan 6, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During product testing of the PLV Continuum Ventilator's Safety Valve/Pressure Relief Valve (SV/PRV) at our manufacturing facility, Respironics found that when the PLY Continuum Ventilator is operated in the 'Face Up' orientation, the SV /PRV may stick in the open position, which may result in loss of delivered gas volume to the patient.

Source facts

Action
On November 18, 2005 all affected customers received a personal phone call from Respironics explaining the recall situation in detail. On November 21, 2005, an advisory notice was sent to customers via fax, email, or mail, instructing customers to only use your PLV Continuum Ventilator in the 'Face Foward' orientation. The firm will contact customers with further information shortly to assure that this situation is positively resolved. Customers were instructed to Only use the PL V Continuum Ventilator in the 'Face Forward' orientation. In the event that a customer requires the application of the PLV Continuum Ventilator in the Face Up Orientation, Respironics will provide the customer, at no charge, with a rental PLV-l00, PLV-IO2b or PLV-I02 Ventilator for the customer to use until their PLV Continuum Ventilator is upgraded. If additional information is needed the customer was instructed to contact Respironics'' Customer Service at 800 345-6443.
Initiated
2005-11-18
Root cause
Other
Terminated
2008-04-22
Distribution
Nationwide, Taiwan, Saudi Arabia, Netherlands, Hong Kong, Canada, Brazil, Australia, and Argentina.
Recall number
Z-0354-06
Recall status
Terminated
CDRH record ID
43173
FDA classified
2006-01-06
Recall event number
34129

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗