Open original FDA record ↗Recall record
- FDA identifier
- Z-0354-2022
- Company in FDA record
- Argon Medical Devices, Inc
- FDA record creation date
- Dec 9, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to component (housing and plunger) detaching from the semi-automatic biopsy instruments.
Source facts
- Action
- On 09/30/2021, the firm sent an "Urgent-Product Recall Notice" letter to customers via US Postal Service Registered Mail informing them that the firm has received complaints regarding the semi-automatic Biopsy instrument coming apart during shipping or prior to use. According to the firm's investigation, the plastic housing and plunger can be separated more easily than normal for the lots manufactured during this time frame. Customers are instructed to return all affected product to the Recalling Firm's Athens TX facility using RGA#26683, attention Andrea Wieczor. Mailing address is: RGA# 26683 Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751 USA Att: Andrea Wieczor In addition, Customers are instructed to complete the response acknowledgement form included with the letter/email and return by fax, mail, or email as soon as possible. For questions. contact the Director, Post-Market Experience at email address Brian.Rogers@argonmedical.com, or contact Arbee Cummings at Arbee.Cummings@argonmedical.com or Andrea Wieczor at Andrea.Wieczor@argonmedical.com. (Updated information 11/2/2021), on 10/28/2021, the firm sent an updated "Urgent-Product Recall Notice for Additional Lots" letter via mail informing customers that they are expanding the recall to additional lots (list on the Inventory Sheet attached to recall notice). Again the Recalling Firm is instructing customer to return the affected product (listed on the Inventory Sheet).
- Initiated
- 2021-10-04
- Root cause
- Device Design
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of ALB, ARE, ARG, AUS, AUT, BEL, BGD, BRA, CAN, CHE, CHL, CHN, COL, CZE, DEU, ECU, ESP, EST, FRA, GBR, GRC, HKG, HUN, IND, IRN, ISR, ITA, JOR, JPN, MAR, MEX, MYS, NLD, NPL, NZL, PAK, PER, PHL, POL, PRT, SAU, SGP, SRB, THA, TUR, and TWN.
- Recall number
- Z-0354-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189610
- FDA record created
- 2021-12-09
- Recall event number
- 88814