Skip to content

Public FDA record

Z-0354-2023: Alcon Custom Pak Surgical Procedure Pack

Recall record Z-0354-2023 for Alcon Research, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0354-2023
Company in FDA record
Alcon Research, LLC
FDA classification date
Dec 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to difficulty removing the liner on the adhesive components of the affected drapes and as a result may render the product unusable. In addition, the manufacturer of the drapes has observed an increase in reported adhesive related skin injuries for these affected lots.

Source facts

Action
On 10/31/2022, the firm sent an "URGENT: VOLUNTARY MEDICAL DEVICE FIELD CORRECTION" Letter via overnight mail informing customers about the problem, the product and the actions to be taken. Customers are instructed to: 1. See table 1 for list of affected Alcon Custom Pak lots (included in list) that were distributed to their facility. They may receive Custom Pak lots assembled prior to this Field Correction notification, and as a result, may be receiving affected inventory in future shipments. If they receive any affected lots in the future, they will be identified by a sticker. The stickers will state (in red print) "URGENT FIELD CORRECTION Discard enclosed 3M surgical drape. Replace with sterile standalone drape." 2. Review inventory of Alcon Custom Pak lots against the attached table. For affected lots of Custom Pak in their inventory, affix the provided stickers directly to the outside of the identified affected Custom Pak units in a location most easily seen for their facility. NOTE: Alcon Custom Pak units and standalone drapes are sterile and should not be opened prior to surgery. 3. For current on-hand affected inventory, upon opening of your identified Alcon Custom Pak for surgical use, remove and dispose of the enclosed 3M Steri-Drape Surgical Drape. Call Alcon Customer Service (1-800-862-5266) in obtaining replacement sterile standalone drapes. 4. For impacted Custom Pak inventory that will ship in the future, Alcon is planning to ship sterile stand-alone substitutions. 5. Keep Table 1 for their records should they receive any affected Alcon Custom Pak lots in the near future. For questions, contact Alcon Customer Service at 1-800-862-5266
Initiated
2022-10-31
Root cause
Nonconforming Material/Component
Distribution
Worldwide Distribution: U.S.(nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and O.U.S. (foreign) to countries of: Australia, Canada, France, Germany, Italy, Mexico, New Zealand, Portugal and United Kingdom.
Recall number
Z-0354-2023
Recall status
Open, Classified
CDRH record ID
196749
FDA classified
2022-12-01
Recall event number
91105

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗