Skip to content

Public FDA record

Z-0355-05: AMPLATZER Delivery System 45-degree curve, order no. 9-DEL-9F-45/80. Product is labeled as Sterile EO.

Recall record Z-0355-05 for AGA Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0355-05
Company in FDA record
AGA Medical Corporation
FDA classification date
Dec 24, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Tubing utilized in the manufacture of the AMPLATZER Delivery and Exchange System sheaths and dilators was in some cases manufactured by our supplier outside of process parameters; causing a residue to be left on the inside of the tubing.

Source facts

Action
Recall notices were sent via e-mail 11/22/04 to physicians, hospitals and distributors who received affected product. Hard copies of the Recall Notices were sent to all customers (international and domestic) via Federal Express 11/22-24/04. A form that is attached to the Recall Notice asks for customers to fill out quantities of affected product and to return the affected product to AGA Medical.
Initiated
2004-11-23
Root cause
Other
Terminated
2006-12-26
Distribution
Product has been distributed throughout the US and Internationally, including Canada, Europe and Australia.
Recall number
Z-0355-05
Recall status
Terminated
CDRH record ID
36274
FDA classified
2004-12-24
Recall event number
30481

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗