Z-0355-2012: SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. SoftPath is a laboratory information system to

Recall record Z-0355-2012 for SCC Soft Computer. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0355-2012
Company in FDA record
SCC Soft Computer
FDA classification date
Dec 2, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

On 06/13/2011, SCC Soft Computer initiated a correction on the following SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. A client reported when they verified the contents of the fax reports, they found one of the specimen diagnoses to be missing. Parts A through D were present as well as Part F, however, Part E diagnosis was missing for one of the patient reports. All affected clients wer

Source facts

Action
SCC Soft Computer initiated a Field Correction on June 13, 2011. All customers with affected software have been notified of this issue. SCC recommends until the corrected patch is loaded to temporarily use one of the following alternative working solutions: 1. Use MS Word Editor functionality within SoftPath application for creating patient reports. In order to ensure Text Control editor is not inadvertently activated, the templates for Text Control editor will be deleted and reinstalled at the clients request by their SCC representative after receipt of the correction. 2. In Report Template set-up, the bottom margin should be made slightly larger than the size of the footer plus its distance from the bottom page border. For further explanation, please consult your SCC representative for further details, or call (727) 789-0100.
Initiated
2011-06-13
Root cause
Device Design
Terminated
2012-05-31
Distribution
Class II Recall - Worldwide Distribution -- USA (nationwide) and the country of Canada.
Recall number
Z-0355-2012
Recall status
Terminated
CDRH record ID
105437
FDA classified
2011-12-02
Recall event number
60451

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗